Gw Instek GDM-8342GPIB Digital-Multimeter, DMM, GDM-8342GPIB Datenbogen

Produktcode
GDM-8342GPIB
Seite von 158
 
 
COMMAND OVERVIEW 
137 
SYSTem:DISPlay 
Turns the Display on/off. 
Parameter: ON | OFF 
Example: SYST:DISP ON 
Turns the display on. 
SYSTem:DISPlay? 
Returns the status of the display 
Return parameter: 0|1, 0=OFF, 1=ON 
SYSTem:SERial? 
Returns the serial number (eight characters/numbers) 
SYSTem:SCPi:MODE 
Sets the SCPI mode.  
Parameter: NORM| COMP  
(NORM=Normal, COMP= Compatible to GDM8246) 
Example: SYST:SCP:MODE NORM 
Sets the SCPI mode to normal. 
SYSTem:SCPi:MODE? 
Returns the SCPI mode.  
Return parameter: NORMAL | COMPATIBLE 
INPut:IMPedance:AUTO 
Sets the input impedance for DCV mode. 
Parameter: ON(10G)|OFF(10M) 
Example: INP:IMP:AUTO ON 
Turns the Automatic input impedance on. 
INPut:IMPedance:AUTO? 
Returns the input impedance mode. 
Return parameter: <Boolean>(0|1) (0=OFF(10M), 1=ON(10G))