For example:  
 
A
+5A
 0V
-5A
C
D
B
E
DC 5A Range
 
A,B: Input exceeds the ADC dynamic range. 
C,D: The DCI offset causes the input to exceed the 
upper ADC dynamic range. 
E: The DCI offset causes the input to exceed the 
lower ADC dynamic range. 
 
The DC current range should be manually selected 
when the following conditions are true:  
 
When DCI measurement is used. 
 
When the signals being measured contain both 
DC and AC components. 
 
When the amplitude of the AC component in 
the measured signal is higher or lower than the 
dynamic range of the range being currently 
selected by the auto-range function. 
Maximum DCI 
Dynamic Range  
  Selected DCV Range 
ADC Dynamic Range 
DC 500μA 
max±0.1mA 
DC 5mA 
max±1mA 
DC 50mA 
max±10mA 
DC 500mA 
max±100mA 
DC 5A 
max±1A 
DC 10A 
max±10A