Testo testo 870-1 , 9 Hz thermography camera, , 160 × 120 pix bolometer matrix 0560 8701 Benutzerhandbuch
Produktcode
0560 8701
5 Performing the measurement
14
5
Performing the measurement
CAUTION
High thermal radiation (e.g. due to sun, fire, furnaces)
Detector may be damaged!
-
Detector may be damaged!
-
Do not point the camera at objects with temperatures exceeding
500 °C.
Ideal framework conditions
•
Building thermography, investigating the building shell:
Considerable temperature difference between inside and outside required
(ideal: ≥ 15 °C / ≥ 27 °F).
(ideal: ≥ 15 °C / ≥ 27 °F).
•
Consistent weather conditions, no intensive sunlight, no precipitation, no
strong wind.
•
To ensure maximum accuracy, the camera requires an adjustment time of
10 minutes after it is switched on.
Important camera settings
•
Emissivity and reflected temperature must be set correctly in order to
precisely determine the temperature. Subsequent adjustment is possible via
the PC software, if required.
the PC software, if required.
•
When auto-scaling is activated, the colour scale is continuously adjusted to
the min./max. values of the current measurement image. This means that
the colour assigned to a specific temperature is changing constantly! To be
able to compare multiple images based on the assigned colour, scaling
must be set manually to fixed values, or must be subsequently adjusted to
uniform values using the PC software.
the colour assigned to a specific temperature is changing constantly! To be
able to compare multiple images based on the assigned colour, scaling
must be set manually to fixed values, or must be subsequently adjusted to
uniform values using the PC software.
5.1
Saving an image
1
- Press
trigger
.
The image is automatically saved.
testo 870-2: regardless of the image
type set, an infrared image is saved
with an attached real image.
type set, an infrared image is saved
with an attached real image.