16
U2300A Series DAQ Service Guide
2
Calibration
Introduction
Prior to checking the performace of the instrument, ensure 
that you have all the equipments listed in the following to 
perform the verification procedure for analog input and 
analog output. 
Equipment Checklist
Agilent U2300A Series USB Multifunction DAQ devices
Agilent Digital Multimeter (DMM) 34401A/34410A
Fluke Calibrator 5520A/5720A
Computer with Agilent IO Library Suites installed
U2901A Terminal Block 
USB Mini- B Cable
U2901A SCSI Cable
Follow the following steps to ensure the instrument gives 
accurate analog input readings and analog outputs. 
Perform the self- calibration procedure. See 
”.
Perform the performance verification procedure. See 
” for information on analog input connection 
 for information on 
analog output connection.
If the instrument gives accurate readings or accurate 
outputs, adjustment procedure is not necessary. If the 
instrument does not give accurate readings or accurate 
outputs, perform the adjustment procedure. See 
”.
After perform the adjustment procedure, repeat the 
self- calibration procedure and then perform the 
verification procedure.
If the instrument still does not give accurate readings or 
accurate outputs, repeat step 3 and step 4.