COMMNAND OVERVIEW 
157 
Events Bits 
OPC: The operation complete bit is set when all 
selected pending operations are complete. This 
bit is set in response to the *OPC command. 
RQC: Not used. 
QUE: The query error bit is set when there is an 
error reading the output queue. This can be 
caused by trying to read the output queue 
when there is no data in it. 
DDE: The Device Dependent Error indicates a 
memory error/lost memory or failure of the 
self-test. 
EXE: The Execution bit indicates an execution 
error due to one of the following:  
-Illegal command parameter 
-Parameter out of range 
-Invalid parameter 
-Command didn’t execute due to an overriding 
operation condition. 
CME: The Command Error bit can be set when 
a syntax error has occurred, an unidentifiable 
header is received or when a <GET> command 
is received within a program message. (Group 
Execute Trigger) as defined in IEEE 488.1. 
URQ: Not used. 
PON: Not used. 
Event Register 
The Event Register will be set to 0 when read. 
Enable Register 
The Enable Register determines which events 
will set the ESB Bit (bit 5) in the Status Byte 
Register.