Gw Instek PEL-3111 Electronic Load PEL-3111 Datenbogen

Produktcode
PEL-3111
Seite von 174
 
 
COMMNAND OVERVIEW 
31 
Query example 
*SRE? 
>12 
Bits 2 and 3 are set in the Service Request Enable 
register. 
*STB? 
Set
Query
 
 
Description 
Reads the Status Byte register. This command will 
not clear the Status Byte register.  
If the Master Summary Status bit (MSS) is set, it 
indicates that there is a reason for a service 
request. 
Query Syntax 
*STB? 
Return parameter  <NR1> (bit weight) 
Bit number / Description 
0/Not used 
 
1/ERR 
 
2/ CSUM 
 
3/ QUES 
 
16 
4/ MAV 
 
32 
5/ ESB 
 
64 
6/ RQS_MSS 
 
128 
7/ OPER 
 
Note: Bit 1 and 2 cannot be set, however bit 2 (ERR) 
can be returned. 
Query example 
*STB? 
>36 
Bits 2 and 5 are set in the Status Byte register. 
*TRG 
Set
 
 
Description 
This command triggers the unit.  
Syntax 
*TRG