PEL-3000 Programming Manual 
58
 
Parameter  
<NRf>[ A ] 
Step resolution.  
 
MINIMUM or MIN 
Minimum step resolution 
 
MAXIMUM or MAX 
Maximum step resolution 
Return parameter  <ASCII string> 
Returns the range and the 
step resolution. 
Example 
:STEP:CCL 0.02mA 
Sets the step resolution to 0.02mA. 
Query example 
:STEP:CCL? 
> CCL:0.00002 
Returns the step resolution (0.00002A). 
[:CONFigure]:STEP:CC 
Set
Query
 
 
Description 
Returns the step resolution for each CC Mode 
range as a string. 
Query Syntax 
[:CONFigure]:STEP:CC? 
Return parameter  <ASCII string> 
Returns the CCH, CCM 
and CCL step resolution 
settings. 
Query example 
:STEP:CC? 
>CCH:0.002, CCM:0.0002, CCL:0.00002 
Returns the CC mode step resolution for each range. 
[:CONFigure]:STEP:CRH 
Set
Query
 
 
Description 
Configures the step resolution for CR High Range. 
Note: The step resolution setting will be 
automatically rounded to the closest multiple of 
the base resolution. 
Syntax 
[:CONFigure]:STEP:CRH {<NRf>[ MHO ]  | MINimum 
| MAXimum } 
Query Syntax 
[:CONFigure]:STEP:CRH? 
Parameter  
<NRf>[ MHO ]  
Step resolution. [MHO]  
 
MINIMUM or MIN 
Minimum step resolution 
 
MAXIMUM or MAX 
Maximum step resolution