Gw Instek PEL-3111 Electronic Load PEL-3111 Datenbogen

Produktcode
PEL-3111
Seite von 174
 
 
COMMNAND OVERVIEW 
65 
External Control Commands 
 
 
[:CONFigure]:EXTernal[:CONTrol] ................................. 65 
[:CONFigure]:EXTernal:LOADonin ................................ 65
 
 
[:CONFigure]:EXTernal[:CONTrol] 
Set
Query
 
 
Description 
Configures the unit for external control or disables 
external control.  
Syntax 
[:CONFigure]:EXTernal[:CONTrol] {OFF | VOLTage | 
RESistance | RINV } 
Related 
Commands 
[:CONFigure]:EXTernal[:CONTrol]? 
Parameter  
OFF 
Disables external control 
VOLTAGE or VOLT 
Sets the unit to external 
voltage control 
 
RESISTANCE or RES 
Sets the unit to external 
resistance control 
 
RINV 
Sets the unit to external 
resistance (inverted) 
control 
Return Parameter  <ASCII string> 
Returns the external 
control configuration  
Example 
:EXTernal VOLT 
Turns external control voltage control on. 
Query example 
:EXTernal? 
>voltage 
The unit uses external voltage control. 
[:CONFigure]:EXTernal:LOADonin 
Set
Query
 
 
Description 
The LOADonin settings determine whether the 
load is turned on when an external switch is closed 
(LOW) or open (HIGH).