Gw Instek PEL-3111 Electronic Load PEL-3111 Datenbogen

Produktcode
PEL-3111
Seite von 174
 
 
PEL-3000 Programming Manual 
92
 
:POWer:VB 
Set
Query
 
 
Description 
Sets or queries the CP mode “B Value” power. This 
command is only applicable to static mode. 
Note: A different “B Value” can be set to different 
current ranges.   
Syntax 
:POWer:VB {<NRf>[ W ] | MINimum | MAXimum } 
Query Syntax 
:POWer:VB? 
Parameter  
<NRf>[W] 
“B Value” power value 
 
MINIMUM or MIN 
Minimum power level 
 
MAXIMUM or MAX 
Maximum power level 
Return parameter  <NR2> 
Returns the “B Value” 
power value. 
Example 
:POWer:VB MIN 
Sets the power value to the minimum. 
Query example 
:POWer:VB? 
>10W 
Returns the “B Value” power setting. 
:POWer:L1 
Set
Query
 
 
Description 
Sets or queries the CP mode “Level1” power. This 
command is applicable to dynamic mode only. 
Note: A different “Level1” value can be set for 
different current ranges.   
Syntax 
:POWer:L1 {<NRf>[ W ] | MINimum | MAXimum } 
Query Syntax 
:POWer:L1? 
Parameter 
<NRf>[ W ] 
“Level1” power value 
 
MINIMUM or MIN 
Minimum power level 
 
MAXIMUM or MAX 
Maximum power level 
Return parameter  <NR2> 
Returns the “Level1” 
power value.