PEL-3000 Programming Manual 
42
 
MINIMUM or MIN 
 
MAXIMUM or MAX 
 
LIMIT or LIM 
Minimum current limit 
value. 
MAXimum current limit 
value. 
Limit the load 
LOFF  
Turn the load off 
Return parameter  Returns a string with OCP setting followed by the OCP 
value. 
Example1 
:OCP LIM 
Sets the OCP setting to limit. 
Example2 
:OCP 77.000 
Sets the OCP value to 77A. 
Query example 
:OCP? 
>LIMIT, 77.000 
The OCP setting is LIMIT and the OCP value is 
77.000A. 
[:CONFigure]:OPP 
Set
Query
 
 
Description 
Sets or queries the OPP trip settings. The OPP limit 
can be set to a specific value or the trip setting can 
be set to either limit the power or to turn the load 
off.  
Syntax 
[:CONFigure]:OPP {<NRf> [W] | MINimum | 
MAXimum | LIMit | LOFF } 
Query Syntax 
[:CONFigure]:OPP?  
Parameter  
<NRf>[W] 
Power limit value. 
MINIMUM or MIN 
 
MAXIMUM or MAX 
 
LIMIT  or LIM 
MINIMUM power limit 
value. 
MAXIMUM power limit 
value. 
Limit the load 
LOFF  
Turn the load off 
Return parameter  Returns a string with OPP setting followed by the OPP 
value.