Gw Instek PEL-3041 Electronic Load PEL-3041 Datenbogen
Produktcode
PEL-3041
OPERATION
95
PROG
01 ~ 16
3. Select a STEP in the selected program.
STEP
01 ~ 16
4. Select Memory and select which memory
location to load for the selected step.
Settings loaded from the memory location will
be used for the selected step.
be used for the selected step.
The same memory location can be used for
multiple steps.
multiple steps.
Memory
M001 ~ M256
5. Set the Run setting for the step.
By default RUN is set to Skip.
The Auto setting will automatically start and go
onto the next step.
onto the next step.
The Manual setting will wait for the user to
press Next[F2] before running the step.
press Next[F2] before running the step.
Run
Skip, Auto, Manual
6. Choose the On-Time in seconds.
The on-time setting determines how long the
load is turned on for the selected step.
load is turned on for the selected step.
The on-time is defined as the total test time
minus the off-time.
minus the off-time.
On-Time
0.1 ~ 60 seconds
7. Choose the Off-Time in seconds.
The off-time setting determines how long the
load is turned off between the end of the current
step and the start of the next step.
load is turned off between the end of the current
step and the start of the next step.
The off-time is defined as the total test time
minus the on-time.
minus the on-time.
Off-Time
Off, 0.1 ~ 60 seconds