OPERATION 
111 
Description 
Fast Sequence configuration is split into Timing 
Edit configuration and Data Edit configuration.  
Timing Edit configuration is used to configure 
all the settings that are common to all the steps 
of the fast sequence. This includes settings such 
as the mode, range, loops and time base. 
Data Edit configuration is used to create the 
actual steps used in each sequence. 
See below for a description of each. 
Timing Edit 
Overview 
A Fast Sequence contains the following timing 
settings for each sequence: 
 
 
 
Setting 
 
Setting Range 
Description 
Memo 
12 characters 
A user-created note for the 
currently selected sequence. 
Mode 
CC, CR 
Operating mode for the 
sequence. 
Range 
ILVL 
Low I range, low V range 
IMVL 
Middle I range, low V range 
IHVL 
High I range, low V range 
ILVH 
Low I range, high V range 
IMVH 
Middle I range, high V range 
IHVH 
High I range, high V range 
Loop 
Infinity,  
01 ~ 9999 
Sets the amount of times to 
loop the selected sequence. 
Last Load 
OFF, ON 
Set the load condition after 
the end of the sequence. 
Last 
0.000000 
The load setting for when 
Last Load is set to ON. 
RPTSTEP 
0001 ~ 1000 
Last step number 
(0001~1000) per loop