OPERATION 
53 
CR Mode  
 
Description 
In Constant Resistance Mode, the unit will 
maintain a constant resistive load by varying 
the current. CR mode uses ohms, Ω (resistance) 
or siemens, S (conductance) for the setting 
units. For more details on CR mode, see the 
appendix on page 187. 
 Warning 
If you change the mode or the range when the load 
is already on, the load will be turned off 
automatically.  
Operation 
1.  Make sure the load is off. 
2.  Press 
Main
3.  Select CR mode with the Mode[F1] soft-key.  
4.  Select the current range with the I Range[F2] 
soft-key.  
 
Range:  
High, Middle, Low 
 
5.  Select the voltage range with the V Range[F3] 
soft-key.  
 
Range:  
High, Low