OPERATION 
67 
Select the Switching Time Configuration for Dynamic Mode 
 
Description 
The switching time for dynamic mode can be 
configured to switch between two preset on-
times (Timer1, Timer2) or by setting a switching 
frequency and duty cycle.  
Operation 
1.  Press 
Main
 > Configure[F5] > Other[F2] and 
set the Dyna. Time setting.  
 
Range:  
T1/T2, Freq. Duty 
Slew Rate 
 
Description 
The current slew rate can be set for CC and CR 
mode. The slew rate setting is used to limit the 
change in current when switching.  
For static mode, only a single slew rate can be 
set. 
Operation 
1.  Make sure the load is off. 
2.  Press 
Main
.  
 
3.  Set the slew rate(s) using the scroll wheel and 
number pad.  
 
For static mode, only a single slew rate can be 
set. 
 
For dynamic mode, set both the rising and 
falling slew rates.  
 
Take the timer settings into consideration when 
setting the slew rates.