PEL-3000 User Manual 
80
 
Protection Settings 
The Protection settings are used to prevent damage to the unit or 
the DUT by excessive current, voltage or power.  
An alarm is generated and a message is displayed on the screen 
when a protection setting is tripped.  When an alarm is activated, 
the load is turned off (or limited), and the ALARM STATUS pin of 
the J1 connector on the rear panel (pin 16) turns on (open collector 
output by a photocoupler). The protection settings can be used 
regardless of whether the remote sense connections are used or not.  
OCP 
 
Description 
For OCP, the PEL-3000 can be configured to 
either limit the current or turn off the load. 
The OCP levels can be set to 10% higher than 
the rating current. 
Operation 
1.  Press 
Main
 > Configure[F5] > Protection[F1] 
and set the OCP Level and OCP Setting.  
 
Range:  
OCP Level: Rating current + 10% 
OCP Setting: LIMIT, Load Off 
Alarm 
 
When OCP Setting is configured to Load Off, a 
message will be displayed on the screen when 
OCP is tripped. The Enter key must be pressed 
to clear the alarm message.  
 
When configured to LIMIT, OCP will be 
displayed on the screen when the OCP is 
tripped and the current will be limited to the 
OCP Level setting.