Fluke 190-102/S 2-channel hand-held oscilloscope, ScopeMeter® 190 series II, hand-held 4025383 Benutzerhandbuch

Produktcode
4025383
Seite von 158
 
Using The Recorder Functions 
 
Recording Scope Waveforms In Deep Memory (Scope Record) 
2
 
  
47 
Using Triggering to Start or Stop Scope Record  
To record an electrical event that causes a fault, it might 
be useful to start or stop recording on a trigger signal: 
Start on trigger to start recording; recording stops when 
the deep memory is full 
Stop on trigger to stop recording. 
Stop when untriggered to continue recording as long as 
a next trigger comes within 1 division in view all mode. 
For the models 190-xx4 the signal on the BNC input that 
has been selected as trigger source must cause the 
trigger. 
For the models 190-xx2 the signal applied to the banana 
jack inputs (
EXT TRIGGER
 
(in)
). signal must cause the 
trigger. The trigger source is automatically set to 
Ext.
 
(external). 
To set up the test tool, continue from step 3 of the 
previous section: 
Apply the signal to be recorded to the BNC 
input(s). 
F1
 
Stop recording to unlock the 
OPTIONS… softkey 
F2
 
Open the 
RECORDER OPTIONS 
menu. 
ENTER
Jump to the Mode: field, select 
on Trigger… (models 190-xx4) or 
on Ext. (models 190-xx2) to open 
the 
START SINGLE SWEEP ON 
TRIGGERING 
or the
 START SINGLE 
SWEEP ON EXT
.
  
menu. 
 
 
ENTER
Select one of the Conditions: 
and accept the selection. 
For external triggering (190-xx2) continue at step 9.