Using The Recorder Functions
Recording Scope Waveforms In Deep Memory (Scope Record)
2
47
Using Triggering to Start or Stop Scope Record
To record an electrical event that causes a fault, it might
be useful to start or stop recording on a trigger signal:
Start on trigger to start recording; recording stops when
the deep memory is full
Stop on trigger to stop recording.
Stop when untriggered to continue recording as long as
a next trigger comes within 1 division in view all mode.
be useful to start or stop recording on a trigger signal:
Start on trigger to start recording; recording stops when
the deep memory is full
Stop on trigger to stop recording.
Stop when untriggered to continue recording as long as
a next trigger comes within 1 division in view all mode.
For the models 190-xx4 the signal on the BNC input that
has been selected as trigger source must cause the
trigger.
has been selected as trigger source must cause the
trigger.
For the models 190-xx2 the signal applied to the banana
jack inputs (
jack inputs (
EXT TRIGGER
(in)
). signal must cause the
trigger. The trigger source is automatically set to
Ext.
(external).
To set up the test tool, continue from step 3 of the
previous section:
To set up the test tool, continue from step 3 of the
previous section:
4
Apply the signal to be recorded to the BNC
input(s).
input(s).
5
F1
Stop recording to unlock the
OPTIONS… softkey
OPTIONS… softkey
6
F2
Open the
RECORDER OPTIONS
menu.
7
ENTER
Jump to the Mode: field, select
on Trigger… (models 190-xx4) or
on Ext. (models 190-xx2) to open
the
on Trigger… (models 190-xx4) or
on Ext. (models 190-xx2) to open
the
START SINGLE SWEEP ON
TRIGGERING
or the
START SINGLE
SWEEP ON EXT
.
menu.
8
ENTER
Select one of the Conditions:
and accept the selection.
and accept the selection.
For external triggering (190-xx2) continue at step 9.