HP Water System HP VXI Benutzerhandbuch

Seite von 34
Programming the Status System  
13
Example commands using Standard Event registers:
*ESE 48
Generate a summary bit on execution or command errors.
*ESE?
Query the state of the Standard Event enable register.
*ESR?
Query the state of the Standard Event event register.
Standard Operation
Status Group
The Standard Operation Status group provides information about the 
state of the measurement systems in an instrument.  This status group is 
accessed through the STATus subsystem.  Standard Operation Status 
includes a condition register, event register, and an enable register.  As a 
beginner, you will rarely need to use this group.  
 illustrates the 
structure of Standard Operation Status.
Figure 2-6. Questionable Data Status Group