Command Reference  
31
*RST;:CAL:ZERO:AUTO 1; :CAL:LFR +60; VAL +0.00000000E+000; 
:DISP:MON:STAT 0; CHAN (@0); :FORM ASC,+7; :FUNC "VOLT"; 
:MEM:VME:ADDR +2097152; SIZE +0; STAT 0; :RES:APER +1.666667E-002; 
OCOM 0; RANG +1.638400E+004; RANG:AUTO 1;:VOLT:APER 
+1.666667E-002; RANG +8.000000E+000; RANG:AUTO 1; :TRIG:COUN +1; 
DEL +0.00000000E+000; DEL:AUTO 1; :TRIG:SOUR IMM; :SAMP:COUN  +1; 
SOUR IMM;TIM +5.000000E-002 S
*OPC
Operation Complete.  Causes an instrument to set bit 0 (Operation Complete 
Message) in the Standard Event Status Register when all pending operations have 
been completed.  By enabling this bit to be reflected in the Status Byte Register 
(*ESE 1 command), you can ensure synchronization between the instrument and an 
external computer or between multiple instruments.
*OPC?
Operation Complete Query.  Causes an instrument to place an ASCII 1 into the 
instrument’s output queue when all pending instrument operations are finished.  By 
requiring the computer to read this response before continuing program execution, 
you can ensure synchronization between one or more instruments and the computer.
*PMC
Purge Macros Command.  Purges all currently defined macros in the selected 
instrument.
*RMC <
name_string>
Remove Individual Macro Command.  Purges an individual macro identified by 
the name_string parameter.
Example
*RMC ‘ LIST’
Remove macro command from *DMC 
example.
*RST
Reset.  Resets an instrument as follows:
Sets the instrument to a known state (usually the power-on state).
Aborts all pending operations.
Disables the *OPC and *OPC? modes.
*RST does not affect:
The state of the VXI interface
The VXI address