Chapter 6. Programming Meter Settings
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
115
6.3  The Battery Charger (cont.)
3.
To condition the batteries, press the 
[ENTER]
 key. The “Condition 
Battery” window should now read “Stop.” The status line should 
change to “Discharging.” (Updating the status could take up to 30 
seconds.)
Note:
It is advisable to condition NiCad batteries when their run time is 
reduced to 50% of their previous run time.
4.
To stop the discharge cycle, press the 
[ENTER]
 key and the “Condition 
Battery” window reappears. The status now changes to “On Charger.”
5.
Press 
[F3]
 (OK) to return to Operate Mode.
6.4 Entering Date and Time 
In Operate mode, the Status Bar displays the current date and time above 
the measurements in the upper right corner of the screen. The Date/Time 
option allows you to set the date or time, which are required for correct data 
logging operation. 
To set the time:
1.
From the Meter menu, scroll to the Date/Time entry and press 
[ENTER]
The Meter Settings window reopens, as shown in Figure 77 on 
page 113. 
2.
From the Units option, press the [
T
] key once to reach the date text 
box. This box displays the current meter date. 
3.
Press 
[ENTER]
 to enter the text window. The meter highlights the center 
number. Use the [
W
] and [
X
] keys to scroll to any number you wish to 
change.