Chapter 6. Programming Meter Settings
126
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
6.10 Resetting Forward and Reverse Totals
On occasion, it might be necessary to clear and reset the forward and 
reverse totals computed by the Forward and Reverse Totalizers. To reset the 
totals:
1.
From the Meter menu, scroll to the Totals entry and press 
[ENTER]
. The 
window now appears similar to Figure 84 below.
Figure 84: Resetting Totals Menu
2.
Use the [
T
] or [
S
] arrow keys to choose the total to be reset.
3.
Press 
[ENTER]
 to reset the total(s).
The meter resets the selected total(s) to 0.0 and returns to Operate Mode.