Chapter 8. Servicing the PT878
166
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
Dual-Point Calibration (cont.)
1.
Press the [
T
] arrow key to enter the window.
2.
The first prompt asks you to choose between single and dual-point
calibration. Use the [
calibration. Use the [
W
] and [
X
] arrow keys to move to the Dual radio
button and press
[ENTER]
.
3.
The next prompt asks you to enter the length of Block 1. (If you are
using a GE-supplied test block, the length is printed on the block.)
using a GE-supplied test block, the length is printed on the block.)
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the length.
c.
Press
[ENTER]
to confirm the entry.
4.
Press [
X
] to move to the Calibrate button, and press
[ENTER]
to
prepare for calibration. Then apply couplant B to the Block 1 surface,
and hold the transducer against the block.
and hold the transducer against the block.
5.
When you have the transducer firmly pressed against the block and the
Ready button appears, press
Ready button appears, press
[ENTER]
again to perform the calibration.
Note:
For single-point calibration, skip to step 9.
6.
The next prompt asks you to enter the length of Block 2.
a.
Use the arrow keys to move to the Block 2 text box.
b.
Press
[ENTER]
to open the text box.
c.
Use the numeric keys to enter the length.
d.
Press
[ENTER]
to confirm the entry.
7.
Press [
X
] to move to the Calibrate button, and press
[ENTER]
to
prepare for calibration. Then apply couplant B to the block surface, and
hold the transducer against the block.
hold the transducer against the block.
8.
When you have the transducer firmly pressed against the block, press
[ENTER] (Ready button) again to perform the calibration.
[ENTER] (Ready button) again to perform the calibration.
9.
Hold the transducer steady until the Set button appears. Press
[ENTER]
to store the calibration.