Chapter 8. Servicing the PT878
170
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
8.4.5 Calculating Velocity (Pipe Material Soundspeed) (cont.)
7.
The “Calculated” box shows the thickness value measured. The PT878 
asks for confirmation of the calculated and current values. Press 
[ENTER]
 to commit the calculated value, or to recalculate the value.
Note:
If you commit the calculated value, the settings in the Material 
window also change from the previous material to “Other” with the 
new soundspeed.
To leave the Velocity window:
Press the [
W
] or [
X
] arrow key to move to another tab.
Press 
[F2]
 (Cancel) to return to Operate Mode without confirming the 
soundspeed value.
Press 
[F3]
 (OK) to confirm the new value and return to Operate 
Mode.
8.5 Programming the Thickness Gauge
While the T-Gauge Display option allows you to program the appropriate 
material and soundspeed, the Thickness Gauge Setup option enables you to 
view or change five parameters:
Low Signal Threshold
Transducer Delay
Signal Inversion
Noise Threshold
Detection Threshold
Note:
The thickness gauge programming settings are entered at the factory. 
You should not change them unless instructed by the factory.
The Programming window appears similar to Figure 115 on page 171.