Chapter 8. Servicing the PT878
190
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
8.10.3 Testing the Watchdog Timer Circuit
The PT878 includes a watchdog timer circuit. If a software error causes the 
meter to stop responding, this circuit automatically resets the meter. A 
properly functioning PT878 restarts if you run the Watchdog Test.
To start the Watchdog Test, scroll to the Watchdog Test option in the Test 
Menu and press 
[ENTER]
. The screen appears similar to Figure 126 below.
Figure 126: Watchdog Test Screen
Press 
[F2]
 (No) to cancel the test and return to the Menu screen, or press 
[F3]
 
(Yes) to start the test. The PT878 should go blank for a few seconds, and 
then restart. If it does not follow this sequence, consult the factory.