Chapter 9. Diagnostics and Troubleshooting
204
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
9.1 Error Code Messages
Error Code messages appear on the screen while you are displaying 
measurements. When logs appear in graph form, errors are indicated by a 
vertical gray bar at the point of error. When they appear in a spreadsheet 
format, errors are indicated by their number, in place of the measurement. 
The Error Code messages are only general descriptions of the possible 
problems. Use Table 7 below to isolate and remedy the problem. If you are 
unable to remedy the problem, contact the factory.
Note:
We suggest that, in order to isolate the indicated problem more 
easily, you obtain a test flowcell. Contact the factory for details.
Table 7: Error Messages
Error
Problem
Possible Cause
Action
E0
 No error.
Displays briefly after dis-
play of another error 
message
None required. 
Measurement is valid.
E1
Low Signal - Poor 
ultrasonic signal 
strength.
Broken cable.
Flowcell problem. 
Transducer problem. 
Electronic failure.
Check transducer cable. See 
Flowcell Problems on page 208 
and Transducer Problems on 
page 210. Check programming 
and transducer spacing. 
Contact the factory if problem 
persists.
E2
Soundspeed Error 
(Soundspeed is 
programmed using 
the Program Menu 
Pipe option, as 
described in 
Chapter 5.)
Poor flow conditions. 
Incorrect programming. 
Bad transducer spacing. 
Measurement is very 
different from the 
programmed 
soundspeed. 
Check soundspeed against 
nominal soundspeed. 
See Flowcell Problems on 
page 208 and Transducer 
Problems on page 210. 
Check programming and 
transducer spacing.