Chapter 3. Programming Site Data
32
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.3  Entering Pipe Parameters (cont.)
d.
If you have selected “Other,” the meter prompts you to enter the 
soundspeed. Use the numeric keys to type the desired soundspeed in 
the text box, and press 
[ENTER]
 to confirm the choice.
Note:
If the “Other” Pipe soundspeed entered is too large, given the 
previously entered Wedge soundspeed and angles, an error message 
similar to Figure 16 below 
will appear. Press 
[F3] (OK) (the error 
message disappears), and enter another soundspeed within the range 
specified.
Figure 16: Soundspeed Error Message Window
Note:
When the corrected soundspeed is entered, a message similar to 
Figure 17 on page 33 appe
ars. Press 
[F3] (OK).to return to the Site 
Menu.
Transducer/Pipe
Cancel
OK
Invalid Value - Too High
100000 is Too High
The valid range is from:
3330.052 to 24000.000 ft/s