Chapter 3. Programming Site Data
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
37
3.5 Entering Fluid Types and Speeds
To access the Fluid option, scroll to the Fluid entry on the Program Menu 
and press 
[ENTER]
. (If you are already in the Transducer/Pipe form, press 
the [X] arrow key to reach the Fluid window, and press 
[ENTER]
.)The 
screen appears similar to Figure 19 below. To step through each parameter, 
press the [T] key. Refer to Figure 145 on page 229 of Appendix A, Menu 
Maps
.
Figure 19: Fluid Type Window
1.
The first prompt asks you to select whether or not you want Tracking 
Windows. These windows are used to detect the receive signal when 
you are unsure of the fluid soundspeed. (Default operation is “No.”)
a.
Use the [
W
] and [
X
] keys to scroll to the appropriate radio button.
b.
Press 
[ENTER]
 to confirm your selection.
         
 Fluid
Cancel
OK
Transducer/Pipe
Transducer
Pipe
Path
Fluid
Lining
Tracking Windows?
No
Yes
Fluid 
Sound Speed
Speed Min
Speed Max
Glycol
Water (2C)
1496
1350
1650
10
m/s
m/s
m/s
%
Temp
°F
77