GE General Electric Oxygen Equipment PT878 Benutzerhandbuch

Seite von 297
Chapter 3. Programming Site Data
42
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.7 Entering the Energy Option Parameters
The Energy Option enables you to calculate the energy of a system based on 
the temperature at a supply point, the temperature at a return point, and the 
flow of fluid through the system. To enter the Energy Option, scroll to the 
Energy entry on the Program Menu and press 
[ENTER]
. The screen appears 
similar to Figure 21 below. To step through each parameter, press the [T] 
key. While programming, refer to Figure 147 on page 231 of Appendix A, 
Menu Maps.
Figure 21: Energy Option Window
1.
The first prompt asks if you want to disable or enable the Energy 
Option.
a.
Use the [
W
] and [
X
] keys to scroll to the appropriate radio button.
b.
Press 
[ENTER]
 to confirm your selection.
Note:
If you select “Disabled,” you cannot select any other prompt in this 
window.
Energy Option
Inputs
Energy Options
OK
Cancel
                 
l
Energy    
Disabled
Enabled
System
Heating
Flow Measurement at
Supply
Return
Cooling
Calculation Method
Standard
Custom Cp
Custom Cp