Chapter 3. Programming Site Data
48
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.7.2 Entering Custom Cp Data in the Energy Option (cont.)
2.
The next prompt asks from which source — a fixed value or table — 
the PT878 will use for fluid density values. Use the [
W
] and [
X
] keys 
to scroll to the appropriate radio button, and press 
[ENTER]
.
3.
The menu now varies, depending on your selection in step 2.
If you selected “Fixed,” the PT878 asks for the fixed fluid density. 
Use the numeric keys to enter the desired value, and press 
[ENTER]
.
If you selected “Table,” the meter highlights the “Edit Table” button. 
Figure 25: Fluid Density Table
a.
Press 
[ENTER]
 to open the Edit Density table, shown in Figure 25 
above.
b.
Use the numeric keys to enter the desired temperature in degrees 
Kelvin, and press 
[ENTER]
.
c.
Press the [
X
] key to move to the kg/m^3 column (density in 
Kilogram/cubic meters). Use the numeric keys to enter the desired 
value, and press 
[ENTER]
.
 Edit Density
Cancel
OK
°Kelvin
kg/m^3
1
2
3
4
5