Chapter 3. Programming Site Data
54
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.9 Entering the Analog Output
The Analog Output option enables you to enter information to set up output 
parameters. To enter the option, scroll to the Analog Output entry on the 
Program Menu and press 
[ENTER]
. The screen appears similar to Figure 29 
below. To step through each parameter, press the [T] key. Refer to 
Figure 149 on page 233 of Appendix A, Menu Maps.
Figure 29: Analog Output Window
1.
The first prompt enables you to select a range to send a current signal to 
a recording device. 
a.
Press 
[ENTER]
 to open the drop-down menu.
b.
Scroll to the desired output from three choices: Off, 0-20 mA, and 
4-20 mA.
c.
Press 
[ENTER]
 to confirm your selection.
Note:
If you select “Off,” you will not be able to access any other 
parameters in this option.
         
Analog Output
Cancel
OK
Function
         Data Source
Zero
On Error
0-20 mA
Velocity
 Span
Hold Last Value
0
Meters/sec
25.5
m/s
m/s