Siemens syngo CT 2005C Benutzerhandbuch
![Siemens](https://files.manualsbrain.com/attachments/fea10bc852adcb1f4f76708d5daebca542e532b9/common/fit/150/50/c53f5738b57089fc3d6c279e23db7450685a19c9616727fc1f659077967b/brand_logo.png)
180
Specials
HeadTrauma
A spiral mode for emergency head studies with a max.
FoV of 500 mm. A typical scan range of 40/80 cm will
be covered in 23/27 sec.
FoV of 500 mm. A typical scan range of 40/80 cm will
be covered in 23/27 sec.
Base
Cerebrum
kV
130
130
Effective mAs
110
110
Rotation time
1.5 sec
1.0 sec
Slice collimation
1.5 mm
2.5 mm
Slice width
3.0 mm
8.0 mm
Feed/Rotation
3.0 mm
5.0 mm
Pitch Factor
1.0
1.0
Increment
3.0 mm
8.0 mm
Kernel
H31s
H31s
CTDI
Vol
25.05 mGy
25.05 mGy
Effective dose
Male:
0.37 mSv
Female:
0.51 mSv
0.51 mSv
Male:
0.70 mSv
Female:
0.86 mSv
0.86 mSv