Siemens syngo CT 2005C Benutzerhandbuch
204
Children
Neck
Indications:
Spiral mode for routine neck studies, e.g., tumors, lym-
phoma, abscesses, etc.
phoma, abscesses, etc.
A range of 15 cm will be covered in 19 sec.
* The conversion factor for a 7-year-old child, and a
scan range of 150 mm was used.
scan range of 150 mm was used.
Neck
kV
130
Effective mAs
26
Rotation time
1.0 sec
Slice collimation
2.5 mm
Slice width
5.0 mm
Feed/Rotation
9.0 mm
Pitch Factor
1.8
Increment
5.0 mm
Kernel
B50s
CTDI
Vol
2.80 mGy
Effective dose
Male: 2.40 mSv*
Female: 2.52 mSv*
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/
Access site
Access site
Total amount
1 – 2 ml per kg of body weight
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 204 Friday, April 8, 2005 9:55 AM