CONNECTING THE UNIT 
CAUTION: A 5-micron filter (TSD800-6), not included, must be installed with the unit to 
ensure proper air filtration. 
 
 
 
Figure 2.0 
 
 
 
 
 
 
 
 
 
SETUP INSTRUCTIONS 
Refer to Figure 1.0  
6.1  Manual Dispense Cycle Setting: 
1.  Turn on the unit by pressing the Power button (1) 
2.  Press the Mode button (2) until “PURGE” appears on the Display 
3.  Turn up the air pressure by rotating the Air pressure regulator knob (6) 
until the desired pressure is indicated on the Air pressure gauge (5).   
4.  After filling the barrel or using pre-packaged Adhesives, attach syringe to 
receiver head assembly.  Make certain that the syringe locks into place. 
5.  Connect the plug end of the receiver head assembly to one of the air 
dispense outlet (7) 
6.  If vacuum “suck back” is needed, rotate the Vacuum regulator (8) counter 
clockwise until the desired vacuum pressure is obtained 
7.  Press and hold the Footswitch to activate dispensed cycle. (The Manual 
Mode is now activated) 
 
6.2   Automatic Dispense Cycle Setting 
1.  Push the Mode button (2) to select “TIMED” mode  
2.  Rotate the Time control knob (9) to set desired dispense time 
3.  Repeat step 3 to 6 in Section 6.1 
4.  Depress Foot switch to activate timed dispensed cycle 
 
 
Items Description 
To Air Source 
Air Filter (Not Included) 
3 Power 
Adapter 
4 Foot 
Switch 
2
1
3
4
 
6
6.3  Cycle Counter 
The cycle counter records the numbers of dispense cycle being activated.  
Up to 65,535 cycles can be recorded.  This number is shown at the lower 
right hand corner of the LCD.  Press the Counter Reset button (3) to clear 
the counter.  
 
TROUBLESHOOTING 
 
PROBLEM POSSIBLE 
CAUSE CORRECTION 
LCD does not 
light 
• No power input 
 
 
• Check power cord 
connections 
• Turn on power 
System will not 
actuate 
• Foot switch not 
plugged in or 
improperly plugged 
in 
• Defective foot switch 
 
• Broken wire or loose 
connection inside 
unit 
 
 
• Defective solenoid 
• Defective PC board 
 
• Check foot switch 
connection 
 
 
• Foot switch needs to be 
repaired or replaced 
• Unplug power cord and 
disconnect air supply.  
Remove cover and check 
for broken wires or loose 
connections 
• Replace solenoid 
• Replace PC board 
 
System will not 
pressurize 
• Insufficient air 
pressure 
• Air hoses not 
plugged in 
• Regulator defective 
• Increase air supply 
pressure 
• Check connection 
 
• Replace regulator 
System will not 
pull vacuum 
• Vacuum setting is 
too low 
• Defective solenoid 
• Defective vacuum 
venturi 
• Defective vacuum 
needle valve 
• Increase vacuum setting 
• Replace solenoid valve 
• Replace venture 
 
• Replace needle valve 
Inconsistent 
dispensing 
• Air bubbles in 
Adhesives 
• Dispense time is too 
low 
• Dispense needle 
started to clog 
• Reduce vacuum setting 
 
• Increase dispensing tim 
 
• Replace needle