National Instruments NI-488.2 Benutzerhandbuch

Seite von 135
Chapter 1
Introduction
1-2
www.natinst.com
Controlling More Than One Interface
Figure 1-2 shows an example of a multiboard system configuration. 
gpib0
 
is the access interface for the voltmeter, and 
gpib1
 is the access interface 
for the plotter and printer. The control functions of the devices 
automatically access their respective interfaces.
Figure 1-2.  Example of Multiboard System Configuration
Configuration Requirements
To achieve the high data transfer rate that the GPIB was designed for, 
you must limit the number of devices on the bus and the physical distance 
between devices. The following restrictions are typical:
A maximum separation of 4 m between any two devices and an 
average separation of 2 m over the entire bus.
A maximum total cable length of 20 m.
A maximum of 15 devices connected to each bus, with at least 
two-thirds powered on.
One
GPIB
Another
GPIB
Digital
Voltometer
Plotter
Printer
gpib0
gpib1