Konica Minolta CM-SA Benutzerhandbuch

Seite von 4
4
6. 
If you were not able to install the USB driver using the procedure above, install it from Device Manager.
From the Start menu, right-click “My Computer” or “Computer” and select “Properties” to open Device 
Manager. Right click “Spectrophotometer CM-700/600 Series” and select “Update Driver”. Select the 
folder where CM-SA has been installed and specify Kmsecm700.inf.
7. 
Check which COM port of the PC the instrument has been assigned to.
From the Start menu, right-click “My Computer” or “Computer” and select “Properties” to open Device 
Manager. Click “Ports (COM & LPT)” to expand it, and the assigned COM port number is displayed.
Note: This COM port number is required when you start CM-SA and connect it with the instrument.
2.2
Connecting to a PC/Establishing Bluetooth
®
 Connection (When the 
Bluetooth
®
 communication is used.)
1. 
Turn on the instrument and the PC.
2. 
Attach the Bluetooth
®
 adapter to the USB port of the PC.
For the installation and operation of the Bluetooth
®
 adapter driver, refer to the instruction manual of the 
Bluetooth
®
 adapter.
3. 
Set the Bluetooth
®
 communication function of the instrument to ON.
For the procedure to set the Bluetooth
®
 communication function, refer to the instruction manual of the 
instrument.
When the Bluetooth
®
 connection is recognized, the instrument will be displayed on the screen of the 
Bluetooth
®
 adapter driver software. The instrument is identified with the device name 
“KMSEA_xxxxxxxx” (xxxxxxxx is the serial number of the instrument) and the device type “PDA”.
4. 
Use the Bluetooth
®
 adapter's driver software to check the COM port of the PC to which the instrument has 
been assigned.
Note: This COM port number is required when you start CM-SA and connect it with the instrument.
3. Major functions of the software 
(For details, refer to the Instruction Manual.)
4. Precautions for measurement
1. 
If there is any opening between the target and measuring head, light leaks in from the opening, resulting in 
a lower Melanin value.
To prevent measurement errors, ensure complete contact of the measuring head as much as possible.
2. 
Be careful because pressing the measuring head too hard against the skin may cause congestion in the skin.
3. 
Naevi (moles) in the dermis (blue naevus, naevus of Ota, etc.) cannot be measured.
4. 
Do not perform measurement with the specimen measuring port directed towards your eyes and closed 
eyelids.
Skin data display
Melanin Index
Hb Index [Hemoglobin (oxidized and reduced) index]
Hb SO
2
 Index (%) [Hemoglobin oxygen saturation index (%)]
Colorimetric value display 
L*, a*, b*, Munsell value (Hue, Value, Chroma)
Graph display
Hue-Value Graph
Hb Index-Melanin Index Graph
Data output
Saving/reading data in CM-SA original format
Saving data in text (CSV) format
[Melanin Index, Hb Index, Hb SO
2
 Index (%), L*, a*, b*, Munsell 
(Hue, Value, Chroma), Spectral reflectance (400-700 nm) ]
Applicable instrument
Spectrophotometer CM-700d/CM-600d/CM-2600d/CM-2500d