Henkel LOCKTITE 97211 Benutzerhandbuch

Seite von 60
20 
 
5    Measuring 
5.3    Setup at the Controller 
   
• Open the main directory PERIPHERY SETUP, see operating manual of the used 
controller. 
   
• Set the directory point Flow.Monitor respectively Disp.Control from [ O ] to [ I ]. The 
subdirectory is open. 
   
• Choose the required tolerance and 
store it. 
Toler.:    low     [ 0 ] / [ 1 ] 
Toler.:    middle [ 1 ] / [ 0 ] 
Toler.:    high    [ 0 ] / [ 1 ] 
Tolerance    35% 
   
The values of these tolerance classes are:     Toler.:   low     15% 
                                                      Toler.:   middle 35% 
                                                      Toler.:   high    55% 
   
Start with the tolerance middle (factory setting). Dispense several times to see, if it is too 
sensitive or too insensitive, e. g. do you get too many error messages due to normal 
fluctuations in the product or no error message at all. 
   
If you get too many error messages, set tolerance to high or increase the tolerance to a 
value, which is big enough that the relationship between ok and bad dispensing 
correspond to your requirements. 
   
If you get too little error messages, set tolerance to low or decrease the tolerance to a 
value, which is small enough that the relationship between ok and bad dispensing 
correspond to your requirements. 
 
At the Automatic Controller 97103 the tolerance values low, middle and high are 
available only.