Airflow Systems IAQ10 Benutzerhandbuch

Seite von 20
 
11 
1. Turn the instrument off. Under the 
batteries, move DIP switch # 7 to ON. 
Turn the instrument back on. 
2. Press and hold the READ key to begin 
the calibration procedure. The display 
will begin a countdown from 5 to 0. 
Release the key when the display reads 
0. If the key is released too soon or too 
late, the instrument will go back to sam-
pling in real-time. “ZerO” will appear 
on the display if the instrument is in 
calibration mode. 
3. Install the regulator on the zero calibra-
tion gas tank and connect tubing from 
the regulator to the inlet fitting on the 
top of the Model IAQ10. The regulator 
has a fixed flow rate of 0.3 L/min. Do 
not use a flow rate lower than 
0.25 L/min or higher than 1.0 L/min.  
4. Turn on gas flow. 
5. Press and release the READ key to begin 
a zero reading. The Model IAQ10 begins 
a 60-second countdown during the zero 
reading. The actual zero reading is 
taken in the last 10 seconds. When the 
countdown is completed, the display