A&D MF-50 Benutzerhandbuch

Seite von 75
MS-70 / MX-50 / MF-50 / ML-50 
16 
 
 
6.3.
6.3.
6.3.
6.3.  Proper Operation for Precision Measurement
Proper Operation for Precision Measurement
Proper Operation for Precision Measurement
Proper Operation for Precision Measurement    
6.3.1.
6.3.1.
6.3.1.
6.3.1.  Operation of the sample
Operation of the sample
Operation of the sample
Operation of the sample    
 
  Use a proper sample quantity. If the quantity is small, precise results may not be possible. 
 
  If the moisture content of the sample (example: plastic) can be estimated that is less 
than 1%, the mass of moisture is not enough for the measurement, when measuring 
light sample, the result may not be accurate. Consider the following sample mass for the 
measurement. 
An estimate of moisture content 
1% 
0.5% 
0.1% 
Necessary mass for measurement 
2 g at least 
5 g at least 
20 g at least 
 
 
  If the measurement is repeated, maintain the same sample quantity. 
 
  Crush grain samples to a small, uniform powder for a quick drying process.   
 
  Spread the sample as evenly as possible. 
 
  The analyzer is designed to measure the moisture content 
of the sample by its weight change. If the sample includes 
volatile matter, it may vaporize during drying causing a measurement error. 
 
  When measuring a liquid or liquid state sample that may make a film on the surface, we 
recommend you use a glass fiber sheet (AX-MX-32-2). 
 
 
6.3.2.
6.3.2.
6.3.2.
6.3.2.  Operation of the analyzer
Operation of the analyzer
Operation of the analyzer
Operation of the analyzer    
 
 Press the 
 RESET 
 key to display the zero value before each 
measurement. 
 
  Check that the displayed sample weight is stable before 
measurement. Press the 
 START 
 key to start a 
measurement. 
 
  Select the proper Analysis Mode to finish a measurement. Use 
the change of moisture content per one minute [%/min] that is 
displayed during measurement as a reference value. 
 
  The analyzer needs a pre-heating process before measurement. When measuring 
samples repeatedly or continuously, the first result is always different from the others. 
 
  The pre-heating process is as follows: Put a sample pan, instead of a sample, on the 
pan. Press the 
 START 
  key to heat it. The analyzer temperature becomes equilibrium. 
 
  Use a sample on the sample pan that has cooled to room temperature. When a sample 
is put on a hot sample pan, the moisture content is diffused before measurement, and 
precise results are not possible. We recommend you use multiple sample pans.   
 
  Do not pile up sample pans during a measurement. 
 
  Avoid drift and vibration of air conditioners. It may cause "measurement error" and 
"unstable value". In particular, because MS-70 is sensitive instrument, it is necessary to 
consider these influences.   
 
  When the difference between ambient temperature and sample temperature is small, it 
may cause temperature control error. 
Example: If the heating temperature is set to 50°C, it is affected by an room temperature. 
 
  Check the activation of the halogen lamp with the operation indicator. 
Example: If low heating temperature is set, the brightness of halogen lamp becomes dark. 
 0.000
g
 
 
 
5.43
%
  2.02
   %/min
 Result display 
No good 
Good 
Check the activation 
with blinking