A&D MF-50 Benutzerhandbuch

Seite von 75
MS-70 / MX-50 / MF-50 / ML-50 
38 
 
 
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display
Set the Minimum Scale Value of the % Display    
  13.  Select the % display with the 
 SELECT 
 key. 
Select 0.01 [%] with the 
 or 
 key. 
 
 
Set the Minimum Scale Value of the Gram Display
Set the Minimum Scale Value of the Gram Display
Set the Minimum Scale Value of the Gram Display
Set the Minimum Scale Value of the Gram Display    
  14.  Select the gram display with the 
 SELECT 
 key. 
Select 0.001 [g] with the 
 or 
 key. 
 
 
 
Store the Parameters and Finish the Operation
Store the Parameters and Finish the Operation
Store the Parameters and Finish the Operation
Store the Parameters and Finish the Operation    
 15.  Press the 
 ENTER 
  key to store the new parameters of 
the measurement program to program number 3. 
Pressing the key, the weighing mode is automatically 
displayed. 
When PROG 3 is recalled, the settings can be used. 
 
 
  To cancel new parameters and return to weighing 
mode, press the 
 RESET 
 key. 
 
 
 
 
Timer Mode
Timer Mode
Timer Mode
Timer Mode    
Refer to page 36 for Standard Mode, Automatic Mode or Manual Mode 
 
 
  This explanation uses the following parameters and sample displays of MX-50. 
 
 
 
 
 
 
 
 
"Temperature 1" can be set higher than "temperature 2" in step drying. 
 
  
Common 
Items 
 
   
Program number.......................................... 6    ( PROG 6 ) 
 
   
Analysis mode ............................................. Timer mode (Symbol:    U-t 
)
 
   
Drying program 
Heating pattern ....................Step drying (
 
 
 
   
 
Drying temperature 1 ..........160°C 
 
   
 
Drying temperature 2 ..........120°C 
 
   
 
Time 1 ..................................5.0 minutes 
 
   
 
Time 2 ..................................10.0 minutes 
  
 Measurement 
unit ...............................................Moisture 
content 
 
   
Minimum scale value during measurement.. 0.01 % 
 
   
Minimum scale value of the gram display .... 0.001 g 
 
   
Sample quantity ........................................... Approximately 5 g 
 
   
Data memory function.................................. Not used 
 
 
Time 1
Drying temperature 2 
Drying temperature 1 
Stage 1 
Stage 2 
Time 2
% MOIST
/W
/D
 0.000
g
 
160°C 
 
Gram display of 
weighing mode 
  
"0."0"1
%
 
   
Minimum scale 
value of % display 
 "0."0"0"1
g
 
   
Minimum scale 
value of gram 
display