13.3.  Calibration of Drying Temperature
Calibration of Drying Temperature
Calibration of Drying Temperature
Calibration of Drying Temperature    (for MS
(for MS
(for MS
(for MS----70 and MX
70 and MX
70 and MX
70 and MX----50)
50)
50)
50)    
 
 
  The temperature calibrator (accessory AX-MX-43) 
adjusts the drying temperature on the pan. Put the 
sensor on the pan and input measurement data at 
100°C and 160°C. 
 
  Each adjustment needs fifteen minutes. The buzzer 
sounds at the end. 
 
 
 t-Up 
 is displayed after no adjustment for five 
minutes during the operation and calibration is 
stopped. Press any key to return to weighing mode. 
 
 Refer to the instruction manual of the certified 
temperature calibrator (accessory AX-MX-43). 
 
 
 
13.3.1.
13.3.1.
13.3.1.
13.3.1.  Operation
Operation
Operation
Operation    
 
  1.  Replace the weighing pan with the sensor pan 
of the temperature calibrator. 
 
  2.  Curve the sensor wire so it does not touch the 
heater cover and glass-housing when closing 
heater cover. 
 
  Level the sensor pan. 
 
  Do not leave a gap between the pan support 
and the sensor. 
 
 
 
 
 
3.  Turn on the analyzer.   
Display the gram unit (of the weighing mode) 
 
 
4.  Press and hold the 
 PROGRAM 
 key  
 
 5. Press 
the 
 SELECT 
  key to display 
 t-Cal 
 
 6. Press 
the 
 ENTER 
 key. 
 
 7. Press 
the 
 START 
  key to start the 100°C 
measurement. 
 
 
  8.  After fifteen minutes, the buzzer sounds and blinks 
 
"1"0"0°C 
. Adjust the blinking value to the 
thermometer value using the 
 or 
 key. (Example: 
97°C) 
 
 
 
 9. Press 
the 
 ENTER 
  key to store the new data and 
to start the 160°C measurement. 
 
15 min.
 0.000
g
 
 t-Cal
 
15 min.
Buzzer 
100℃  "1"0"0℃
 
100℃  "9"7℃
 
Leave no gap 
Leave no gap 
Level the sensor pan. 
Curve the sensor wire so as not to touch heater cover. 
Leave no gap 
Sensor pan
Thermometer 
Sensor pan