Chapter 3   Admit and begin monitoring
Welch Allyn  Acuity Central Monitoring System
To respond to the CONFIRM PATIENT? prompt 
1.
If you see the CONFIRM PATIENT? prompt, respond as follows:
If the indicated patient is the correct patient, press YES
The patient is confirmed at the Acuity Central Station.
The Acuity System sends unit default alarm limits and unit settings to the patient 
monitor.
If the indicated patient is incorrect, press NO.
The MONITOR SETTINGS? prompt appears. 
2.
If you see the MONITOR SETTINGS? prompt, click one of these buttons:
.
At the central station, the patient map icon remains in the yellow ID or RM 
unconfirmed state until the ID and room are confirmed.
WARNING   Carefully verify the patient’s ID and name before you respond to the 
CONFIRM PATIENT? prompt. If you assign another patient’s ID number, the 
Acuity System could combine patients’ Full Disclosure review information.
NEITHER
PRESETS
BOTH
CLR TREND
Clears current trend 
values. 
Uses Acuity System 
Presets-- default alarm 
limits and displays.
Keeps current trend 
values. 
Keeps current alarm 
limits and displays.
Clears current trend 
values.
Keeps current alarm 
limits and displays.
Keeps current trend 
values.
Uses Acuity System 
Presets-- default alarm 
limits and displays.