GE ISOVOLT 3003 Bedienungsanleitung

Seite von 4
GE Inspection Technologies
Xray
GEInspectionTechnologies.com
Technical Data (2)
Exposure Time
The equipment has one programmable timer with a non-volatile memory.
Preselection and setting      
 
digital setting from 0.1 to 99.9 minutes
Setpoint value display      
 
digital, 3 digits
Actual value display   
 
 
digital, 3 digits; the remaining time is displayed,  
 
 
 
 
 
i.e. after a mains failure exposure can be continued without any time error.
Prewarning Time
Preselection and setting      
 
digital setting from 2 to 250 seconds or de-activated
Programmed Mode
Number of storable programs    
 
250
Tube Parameters (Limit Data)
Number of memory locations  
for  tube parameters  
 
 
8  
 
 
 
 
 
with digitally assigned tube limit data, e.g. nominal voltage,  
 
 
 
 
 
limit filament current, anode power dissipation
Technology
20 kHz IGBT technology for high voltage circuit and filament current circuit
Emitted interference 
 
 
as per EN 55011A, Class A
Immunity to interference   
 
as per IEC 801-2/1991, 801-3/1984, 801-4/1988
Connected Loads and Mechanical Dimensions
Power connection  
 
 
1/N~ 230 V ±10%, 50/60 Hz
Power consumption 
 
 
4150 W at 60 kV, 58 mA
Max. mains current 
 
 
24 A
Mains fuse 
 
 
 
32 A (as per VDE)
Ground terminal   
 
 
6 mm² Cu
Type of protection               
 
as per EN 60529, IP00 (rack mount), IP32 (desk-top) 
Cooling water connection   
 
3/4 inch
Cooling water consumption  
 
minimum 3.5 l/min, minimum 4.5 bar, maximum 7 bar, 
 
 
 
 
 
at outlet point pressure-free
Cooling water temperature   
 
> dew point < 25° C
Dimensions overall 
 
 
483 mm x 266 mm x 680 mm (WxHxD), 19“ (48 cm) rack mount
Weight   
 
 
 
approximately 57 kg
Specifications are subject to change without notice.
©2004 General Electric Company. All rights reserved.