How do I Verify that the Test Set is Operating Properly?
How do I Verify that the Test Set is Operating 
Properly? 
If your Test Set powers-up and displays the CDMA ANALYZER screen, 
but you suspect an instrument problem, use the Instrument Quick 
Check to verify operation of the basic instrument functions. 
If no failure is indicated by this test, but you still suspect a problem, 
refer to the “Performance Tests” information in the Assembly Level 
Repair Manual
Instrument Quick Check
Step 1.  Set up the quick check:
a. Remove any connected cables (except for AC or DC power). 
b. Turn instrument power on (if it is not already on). 
c. Press 
Preset
.
d. Press 
Inst Config
 and set the 
RF Display
 field to 
Freq
.
e. Press 
CDMA Anl
 and select the 
Avg Pwr
 field, then select 
Rho
 
from the Choices: menu. 
f. Set the 
Tune Freq
 field to 800 MHz.
g. Set the 
Anl Dir
 field to 
Rev
.
h. Press 
CDMA Gen
 and set the 
Output Port
 field to 
RF Out
i. Set the 
RF Gen Freq
 field to 800 MHz. 
j. Set the 
Amplitude
 field to 
40 dBm
k. Set the 
Gen Dir
. field to 
Fwd
. (The 
EQ In/Out
 field is 
displayed.)
Step 2. Check the following reading:
a. Rho should be greater than 0.96.
Step 3. Press 
Spec Anl
 to access the SPEC ANL screen. 
You should see the characteristic CDMA spectrum.
Main Menu