Energizer 515 Benutzerhandbuch

Seite von 188
104 Chapter: 6
3
In the 
SET TEMP BASAL 
screen, the temporary 
basal rate will flash. Enter your temp basal rate, 
then press 
ACT
.
 
4
The BASAL MENU will appear. Your 
temp basal is now set and delivering. 
Exit the menus. 
Verifying temp basal delivery
Temporary basal information is 
available in the STATUS screen only.
During a temporary (temp) basal, 
the pump is in Special mode (an open 
circle appears). This open circle will 
remind you that a temp basal is active. 
Additionally, your pump will 
beep/vibrate three times every hour 
during delivery. During delivery, the 
STATUS screen will show the current 
temp basal information.
appears if 
Temp Basal Type is set 
to “Insulin rate (U/H)”
appears if
Temp Basal Type is set 
to “Percent of basal” 
SET TEMP BASAL U/H
3:00P
0.05
SET TEMP BASAL%
3:00P
60%
OR
U
H
pump is in Special mode 
during a temp basal
BASAL MENU
Cancel Temp Basal 
Select Patterns
Set/Edit Temp Basal
indicates a U/H temp basal of
0.6 units per hour is now active
regular programmed
basal rate that will resume after the
temp basal delivery is finished
temp basal delivery
will last 30 minutes
time remaining before temp
basal delivery is finished
STATUS
 U100 
Tmp Basal 0.60 U/H
Duration 0:30
Time left 0:22
Last Bolus N 3.8U
9:07A 11 JAN
Basal 1: 0.80 U/H
Reservoir Started:
09JAN, 6:03P
Units left: 128U
Time left: 24 Hours
Battery: Normal
Tue 11 JAN 2004
S/N# 000000