Fluke Lawn Mower 2635A Benutzerhandbuch

Seite von 268
2635A
Users Manual
A-14
Maximum Autoranging Time
See Table A-18 (shown in seconds per channel).
Totalizing Input
Input Voltage
30V maximum
4V minimum
2V peak minimum signal
Isolation
None
dc-coupled
Table A-18. Autoranging Rates
Function
Range Change
Slow
Fast
VDC
300 mV
----------------------------
>
150 V
0.25
0.19
150V
----------------------------
>
300 mV
0.25
0.18
VAC
300 mV
----------------------------
>
150V
1.40
1.10
150 mV
----------------------------
>
300 mV
1.40
1.10
Ohms
300 
Ω
----------------------------
>
10.0 M
Ω
1.70
0.75
10.0 M
Ω
----------------------------
>
300 
Ω
1.50
0.60
Threshold
1.4V
Hysteresis
500 mV
Input Debouncing
None or 1.75 ms
Rate
0 to 5 kHz with debouncing off
Maximum Count
65,535
Digital Inputs
The specifications for the digital inputs are provided in the following paragraphs.