Fluke Lawn Mower 2635A Benutzerhandbuch

Seite von 268
2635A
Users Manual
2-30
Using the Review Mode
Perform the procedure in Figure 2-18 to operate the Review Mode of operation. The
Review Mode is used any time during or after operation of the Scan Mode. While the
instrument is in the Scan Mode, the last, maximum, and measurements for each scanned
channel are stored in memory and updated with each scan, forming the Review Array.
For example, during scan operations, 
N can be used to monitor the maximum
measurement of a channel in real time. The Review Array is cleared by a control
sequence (see Figure 2-18 below), or by changing any parameter of any channel or the
measurement rate. The Review Array can be printed out using the 
L key (see Figure
5-4). If the instrument is in the Review Mode and scanning using the memory card, any
illegal memory card operations is noted only with a double "beep." If a double beep is
heard, exit the Review Mode and check the memory card error (see Figure 2-16).
REVIEW
CH
REVIEW
REVIEW
LAST
MIN
MAX
REVIEW
Examining the Review Values. Press the 
REVIEW key to start the Review Mode.  Use the 
up/down arrow keys to select the channel for 
review, then use the left/right arrow keys to view 
the LAST (Last), MIN (Minimum), and MAX 
(Maximum) values recorded during the past scan 
intervals (scan can be active or inactive).  Repeat 
for each channel of interest.  A series of dashes, 
- - - - -, indicates all review data has been cleared 
either by a control sequence (below), or by 
changing any parameter of any channel.  OL 
indicates an overload.  Press the REVIEW key to 
exit the Review Mode.  If not cleared, review 
values will remain in memory for update with the 
next scan interval.
REVIEW
REVIEW
SHIFT
REVIEW
Clearing the Review Array.  Press the Review 
key to start the Review Mode.  Press the SHIFT 
key and then the REVIEW key to clear all review 
data from all channels.  If the scan mode is not 
active, the display will change to dashes (- - - - -).  
If the scan mode is active, new values will 
appear.  Press the REVIEW key again to exit the 
Review Mode.
REVIEW
- - - - -
op30f.eps
Figure 2-18. Using the Review Mode