Multi-Echo EPI 
Page 3 of 9 
Numaris/4 VA21B 
1  Basic principles  
 
 
1.1  Principle of Multi-Echo EPI 
 
The basic timing scheme of Multi-Echo EPI is shown in Figure 1. It is based on the 
original EPI sequence proposed by Mansfield (1). After a single RF-excitation, n 
complete EPI images are acquired in a single shot. The phase gradient is rewound to 
the original starting position after each echo-image to ensure identical k-space 
trajectories for all echo-images (2). 
G
P
RF
G
S
G
R
echo-image loop
n
 
 
Figure 1 – Basic timing of the Multi-Echo EPI sequence. The shaded gradient in 
phase encoding direction ensures identical k-space trajectories for all echo-images. 
 
 
The resulting echo-images have different increasing echo times and thus T2*-
weighting. The single echo-images correspond to standard gradient echo EPI images 
with the same echo time.