GENERAL SAFETY
This equipment has been designed to minimize exposure of personnel to hazards .
While the Polaris System is designed for safe operation, certain precautions must
be observed . Use of the Polaris System not in conformance with the instructions
specified in this manual may result in permanent failure of the unit .
KNOWLEDGE OF WARNINGS AND CAUTIONS
Users  must  exercise  every  precaution  to  ensure  personnel  safety,  and  be
familiar  with  the  warnings  and  cautions  presented  throughout  this  manual
and summarized below .
Warnings.
 To prevent fire or electrical shock, do not expose this equipment to rain or
moisture .
 Do  not  spray  cleaning  liquids  or  disinfectants  directly  on  the  camera
handpiece or the Handpiece Cable . Use care not to allow liquids to run
into internal circuitry .
 Do not wipe the surfaces using benzine, thinner, etc . as this may degrade
the finish .
Cautions.
 There  are  no  user-serviceable  parts  inside .  Servicing  should  be  per-
formed by qualified dealer service representatives only .
 Connecting any device to the Polaris System that does not meet the equivalent
safety  requirements  of  the  system  may  reduce  the  safety  effectiveness  of  the
Polaris System .
 Unplug the system components from power before performing cleaning
procedures .
Contraindications.
None known .
Do Not Attempt Internal Service.
The interior of the camera is only accessible by removing hardware with tools
and should only be opened and serviced by an authorized service technician .
Contact your local Air Techniques authorized dealer for service . Failure to heed
this directive may result in equipment damage and voiding the warranty .
Use of Accessory Equipment.
All peripheral equipment (computers, monitors, printers, etc) connected to this
equipment must comply with IEC 60601-1 and the resulting system must comply
with IEC 60601-1 and/or IEC 60601-1-1 harmonized national standards when
set-up within the immediate patient vicinity (ie . within 6 feet) .
Accessory equipment outside the patient vicinity must comply with IEC 60601-1
or IEC 60950-1 as allowed by IEC 60601-1-1 .
Use of ACCESSORIES or cables other than those specified or provided by Air
Techniques may result in increased EMISSIONS or decreased IMMUNITY of the
equipment .