9. Remote Operation
9.1 Remote Interface
The calibrator can be remotely controlled using a PC terminal, or by
a computer program running the calibrator in an automated system.
It uses an RS-232 serial port connection for remote operation.
NOTE: To use the remote control option a custom RS-232
cable must be purchased from OMEGA (LEM232). To contact
OMEGA refer to Section 1.1 of this manual. With this connection
the user can write programs on a PC, with Windows languages like
Visual Basic to operate the calibrator, or use a Windows terminal,
such as Hyper Terminal, to enter single commands. Typical RS-232
remote configurations are shown in Figure 13.
a computer program running the calibrator in an automated system.
It uses an RS-232 serial port connection for remote operation.
NOTE: To use the remote control option a custom RS-232
cable must be purchased from OMEGA (LEM232). To contact
OMEGA refer to Section 1.1 of this manual. With this connection
the user can write programs on a PC, with Windows languages like
Visual Basic to operate the calibrator, or use a Windows terminal,
such as Hyper Terminal, to enter single commands. Typical RS-232
remote configurations are shown in Figure 13.
Figure 13. Calibrator-to-Computer Connection
9.2 Setting up the RS-232 Port for Remote Control
Note: The RS-232 connection cable should not exceed 15m unless
the load capacitance measured at connection points is less than
2500pF.
the load capacitance measured at connection points is less than
2500pF.
Serial parameter values:
9600 baud
8 data bits
1 stop bit
no parity
Xon/Xoff
EOL (End of Line) character or CR (Carriage Return) or both