6-8 Calibration
ChartScan User’s Manual
Gain Calibration of Main Unit
Required Equipment*
- Volts DC Calibrator
- 22-2 Twisted pairs (to connect calibrator to calibration card)
- Digital Voltmeter, or Digital Multimeter
- 22-2 Twisted pairs (to connect calibrator to calibration card)
- Digital Voltmeter, or Digital Multimeter
(used for verification of calibration voltages)
- Calibration Card (installed in bottom slot, slot #1, of main unit)
*
see required specifications on page 6-2
CAUTION
CAUTION
CAUTION
CAUTION
Offset calibration of main unit must be completed prior to calibrating the main unit for
gain.
gain.
CAUTION
CAUTION
CAUTION
CAUTION
Use approved ESD precautions, including static-free work area and grounded wrist strap,
when handling circuit boards and electronic components. Failure to do so could cause
equipment damage due to electrostatic discharge.
equipment damage due to electrostatic discharge.