6.8.3
Control Register 3 Tab
6.8.4
Control Register 4 Tab
www.ti.com
Kit Operation
Figure 8. Control Register 2 Tab
The Control Register 3 Tab (
allows the user to set the DAC Oversampling Rate (OSR), set
the Asynchronous Sampling Rate Factor and power down the ADC or DAC with the Power Down
Controls
. The indicator below each power down switch will light when on only if the register data sent by
pressing the Program Device button is acknowledged. The USB-MODEVM Audio Interface
Configuration 
currently allows a fixed sampling rate and a single codec. For an OSR=256 the value of M,
set in register 4, must be a multiple of 2. Similarly, for an OSR=512 the value of M must be a multiple of 4.
Figure 9. Control Register 3 Tab
The Control Register 4 Tab (
provides controls for Pand M. Furthermore, an FS calculator
is provided for convenience. The calculator derives FS from the MCLK frequency entered by the user or
loaded by a preset by using the equation: FS=MCLK
÷
(16
×
P
×
M
×
N). The PLL Method switch illustrates
that for coarse sampling, must be equal to 8. Please note that the FS calculator and the PLL Method